Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
881
User Manual 1178.8520.02 ─ 08
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>[:RELative]:STATe
<State>[,
<State>]
This command turns the relative limit check for the alternate channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe]
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<ch>
1..n
Alternate channel number
Parameters:
<State>
ON | OFF | 1 | 0
Relative limit check for lower alternate channel
*RST:
0
<State>
ON | OFF | 1 | 0
Relative limit check for upper alternate channel
*RST:
0
Manual operation:
See
See
CALCulate<n>:LIMit<li>:ACPower[:STATe]
<State>
This command turns the limit check for ACLR measurements on and off.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
See
See
12.5.3.6
General ACLR measurement settings
The following commands control the measurement algorithm.
Useful commands for the ACLR measurement described elsewhere:
●
●
●
[SENSe:]POWer:ACHannel:PRESet:RLEVel
Configuring and performing measurements