Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1037
User Manual 1178.8520.02 ─ 08
INPut<ip>:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
for the R&S
FSV/A after an overload condition occurred and the protection mechanism
intervened. The error status bit (bit 3 in the
STAT:QUES:POW
status register) and the
INPUT OVLD
message in the status bar are cleared.
(See
STATus:QUEStionable:POWer[:EVENt]?
ter 11.2.2.12, "STATus:QUEStionable:POWer register"
The command works only if the overload condition has been eliminated first.
For details on the protection mechanism, see
Suffix:
<ip>
.
1 | 2
irrelevant
Example:
INP:ATT:PROT:RES
INPut<ip>:CONNector
<ConnType>
Determines which connector the input for the measurement is taken from.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<ConnType>
RF
RF input connector
RFPRobe
Active RF probe
*RST:
RF
Example:
INP:CONN RF
Selects input from the RF input connector.
Manual operation:
See
INPut<ip>:COUPling
<CouplingType>
This command selects the coupling type of the RF input.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<CouplingType>
AC | DC
AC
AC coupling