Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1249
User Manual 1178.8520.02 ─ 08
CALC:LIM3:TRAC2:CHEC ON
//Activates the upper limit to be checked against trace2 (average trace)
CALC:LIM:CLE
//Clears the previous limit check results
//------------- Performing the measurement---------------------
INIT;*WAI
//Initiates a new measurement and waits until the last sweep has finished.
//-------------- Retrieving limit check results----------------------------
CALC:LIM1:FAIL?
//Queries the result of the upper limit line check
CALC:LIM3:FAIL?
//Queries the result of the lower limit line check
12.9
Managing settings and results
The commands required to store and load instrument settings and import and export
measurement results in a remote environment are described here.
The tasks for manual operation are described in
Addressing drives
The various drives can be addressed via the "mass storage instrument specifier"
<msis> using the conventional Windows syntax. The internal hard disk is addressed by
"C:".
For details on storage locations refer to
Chapter 9.3.2.2, "Storage location and file-
The file names (<FileName> parameter) are given as string parameters enclosed in
quotation marks. They also comply with Windows conventions. Windows file names do
not distinguish between uppercase and lowercase notation.
Wildcards
The two characters "*" and "?" can be used as "wildcards", i.e., they are variables for a
selection of several files. The question mark "?" replaces exactly one character, the
asterisk replaces any of the remaining characters in the file name. "*.*" thus means all
files in a directory.
Path names
Storage locations can be specified either as absolute (including the entire path) or rela-
tive paths (including only subfolders of the current folder). Use the
MMEM:CDIR?
query
to determine the current folder.
Managing settings and results