![Rohde & Schwarz FSV3000 User Manual Download Page 885](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_1477843885.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
885
User Manual 1178.8520.02 ─ 08
Note that the position of the first Tx channel in the first sub block and the last Tx chan-
nel in the last sub block also affect the position of the adjacent channels.
This command is for MSR signals only (see
Suffix:
<sb>
.
1 to 8
sub block number
<ch>
1..n
Tx channel number
Parameters:
<Frequency>
absolute frequency in Hz
Default unit: Hz
Manual operation:
See
[SENSe:]POWer:ACHannel:SBLock<sb>:FREQuency:CENTer
<Frequency>
This command defines the center of the specified MSR sub block. Note that the posi-
tion of the sub block also affects the position of the adjacent gap (CACLR) channels.
This command is for MSR signals only (see
Suffix:
<sb>
.
1 to 8
sub block number
Parameters:
<Frequency>
absolute frequency in Hz
Default unit: Hz
Manual operation:
See
MSR adjacent channel setup
The functions for manual operation are described in
Chapter 5.3.5, "MSR ACLR config-
Useful commands for MSR adjacent channel setup described elsewhere:
●
Chapter 12.5.3.5, "Checking limits"
●
Remote commands exclusive to MSR adjacent channel setup:
[SENSe:]POWer:ACHannel:SBLock<sb>:RFBWidth
..........................................................886
[SENSe:]POWer:ACHannel:SBLock<sb>:TECHnology[:CHANnel<ch>]
...............................886
[SENSe:]POWer:ACHannel:SBLock<sb>:TXCHannel:COUNt
............................................ 887
[SENSe:]POWer:ACHannel:SPACing:UACHannel
.............................................................887
[SENSe:]POWer:ACHannel:SPACing:UALTernate<ch>
......................................................888
................................................................................888
Configuring and performing measurements