Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1139
User Manual 1178.8520.02 ─ 08
[SENSe:]ADJust:CONFigure:TRIGger
<State>
Defines the behavior of the measurement when adjusting a setting automatically (using
SENS:ADJ:LEV ON
, for example).
See
"Adjusting settings automatically during triggered measurements"
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
[SENSe:]ADJust:FREQuency
This command sets the center frequency to the frequency with the highest signal level
in the current frequency range.
Example:
ADJ:FREQ
Manual operation:
See
"Adjusting the Center Frequency Automatically (Auto Fre-
[SENSe:]ADJust:LEVel
Initiates a single (internal) measurement that evaluates and sets the Thus, the settings
of the RF attenuation and the reference level are optimized for the signal level. The
R&S
FSV/A is not overloaded and the dynamic range is not limited by an S/N ratio that
is too small.
Example:
ADJ:LEV
Usage:
Event
Manual operation:
See
"Setting the Reference Level Automatically (Auto Level)"
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:AUTO:LIST
<List>
Defines the standard to be loaded for an automatic measurement, that is: after
The query returns the list of available standards for the measurement.
This command is only available for ACLR and SEM measurements.
Suffix:
<n>
.
<m>
<sb>
irrelevant
Setting basic measurement parameters