![Rohde & Schwarz FSV3000 User Manual Download Page 1345](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_14778431345.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1345
User Manual 1178.8520.02 ─ 08
OCCupy
Occupied license
<Days>
The remaining time period an occupied license is available.
Range:
1 to 7
Default unit: days
Example:
Query which type of license is available for K100:
SYST:OPT:LIC? 'K100'
//Result: K100 floating
Example:
Query which licenses are available for the R&S
FSV/A:
SYST:OPT:LIC?
//Result:
B4,B5,B21,B114,K10 disabled,K91 floating,K91AC
floating,K91N disabled,K100 occupied
floating,K101 floating,K102 floating,K103
floating,K104 floating,K105 floating,K106
floating
Usage:
Query only
Manual operation:
See
SYSTem:SECurity[:STATe]
<State>
Activates or queries secure user mode.
Note:
Before you activate secure user mode, store any instrument settings that are
required beyond the current session, such as predefined instrument settings, trans-
ducer files, or self-alignment data.
Note:
Initially after installation of the R&S
FSV/A-K33 option, secure user mode must
be enabled manually once before remote control is possible. This is necessary to
prompt for a change of passwords.
For details on the secure user mode see
Chapter 3.2.15, "Protecting data using the
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
The R&S
FSV/A automatically reboots and starts in secure user
mode. In secure user mode, no data is written to the instru-
ment's internal solid-state drive. Data that the R&S
FSV/A nor-
mally stores on the solid-state drive is redirected to SDRAM.
OFF | 0
The R&S
FSV/A is set to normal instrument mode. Data is
stored to the internal solid-state drive.
Note: this parameter is for query only. Secure user mode cannot
be deactivated via remote operation.
*RST:
0
Configuring the R&S
FSV/A