![Rohde & Schwarz FSV3000 User Manual Download Page 941](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_1477843941.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
941
User Manual 1178.8520.02 ─ 08
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Function>
OFF | MAX
Defines the function to be used to determine the relative limit
line stop value
MAX
The maximum of the relative
and
the absolute level is used as
RANGe<ri>:LIMit<li>:RELative:STOP
]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:
STOP:ABS
commands to define these values.
OFF
No function is used, the relative limit line is defined by a fixed
relative stop value. Use the
RANGe<ri>:LIMit<li>:RELative:STOP
this value.
*RST:
OFF
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP:
FUNCtion MAX
Chapter 12.5.6.11, "Example: SEM
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:STATe
<State>
This command selects the limit check mode for
all
SEM ranges (<range> is irrelevant).
Suffix:
<sb>
.
1 to 8
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<State>
ABSolute | RELative | AND | OR
ABSolute
Checks only the absolute limits defined.
RELative
Checks only the relative limits. Relative limits are defined as rel-
ative to the measured power in the reference range.
Configuring and performing measurements