Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
873
User Manual 1178.8520.02 ─ 08
Manual operation:
See
See
[SENSe:]POWer:ACHannel:FILTer:ALPHa:CHANnel<ch>
<Alpha>
This command defines the roll-off factor for the transmission channel weighting filter.
Suffix:
<ch>
.
1..n
Tx channel number
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Manual operation:
See
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ACHannel
<State>
This command turns the weighting filter for the adjacent channel on and off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
See
[SENSe:]POWer:ACHannel:FILTer[:STATe][:ALL]
<State>
This command turns the weighting filters for all channels on and off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALTernate<ch>
<State>
This command turns the weighting filter for an alternate channel on and off.
Suffix:
<ch>
.
1..n
Alternate channel number
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
See
Configuring and performing measurements