Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
416
User Manual 1178.8520.02 ─ 08
File Name
Defines the name under which the table is stored in the
C:\R_S\INSTR\USER\cvl\
directory on the instrument. The name of the table is identical to the name of the file
(without extension) in which the table is stored. This setting is mandatory. The
.ACL
extension is automatically appended during storage.
Remote command:
Comment
An optional comment that describes the conversion loss table. The comment is user-
definable.
Remote command:
[SENSe:]CORRection:CVL:COMMent
Band
The waveguide or user-defined band to which the table applies. This setting is checked
against the current mixer setting before the table can be assigned to the range.
For a definition of the frequency range for the pre-defined bands, see
.
Remote command:
Harmonic Order
The harmonic order of the range to which the table applies. This setting is checked
against the current mixer setting before the table can be assigned to the range.
Remote command:
[SENSe:]CORRection:CVL:HARMonic
Bias
The bias current which is required to set the mixer to its optimum operating point. It
corresponds to the short-circuit current. The bias current can range from -10
mA to
10
mA. The actual bias current is lower because of the forward voltage of the mixer
diode(s).
Tip:
You can also define the bias interactively while a preview of the trace with the
changed setting is displayed, see
Remote command:
Data input and output