General instrument setup
R&S
®
FSVA3000/ R&S
®
FSV3000
734
User Manual 1178.8520.02 ─ 08
Figure 10-2: Preview of selected user correction file
Remote command:
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:DATA:FREQuency?
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:DATA:MAGNitude?
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:DATA:PHASe?
[SENSe:]CORRection:FRESponse<si>:USER:SLISt<sli>:DATA:
FREQuency<spi>?
[SENSe:]CORRection:FRESponse<si>:USER:SLISt<sli>:DATA:
MAGNitude<spi>?
[SENSe:]CORRection:FRESponse<si>:USER:SLISt<sli>:DATA:
PHASe<spi>?
IQ Mode ← Preview
The preview in IQ Mode indicates the frequency ranges covered by the individual cor-
rection files. The blue bar indicates the valid frequency range for which all files contain
values.
An LED-like symbol indicates whether each file is active and user correction in general
is switched on (symbol "lights up") or not (symbol is dark). A purple symbol represents
a Touchstone file. A green symbol represents a frequency response (
.fres
) file.
The lower part shows the combined correction values for the valid frequency range.
Frequency response correction (R&S
FSV/A-K544)