
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1319
User Manual 1178.8520.02 ─ 08
12.10.4.2
Programming example: using touchstone files
This example demonstrates how to use Touchstone files in a basic I/Q Analyzer mea-
surement setup in a remote environment.
-----------Configuring the measurement ------------
*RST; *WAI
SYST:DISP:UPD ON
TRACe1:IQ ON
TRACe:IQ:EVAL ON
LAY:REPL:WIND "1",FREQ
INIT:CONT OFF
// Use calibration signal as input data
DIAG:SERV:INP CAL
DIAG:SERV:INP:RF:SPEC BRO
DIAG:SERV:INP:PULS:CFR 1MHz
// Set SR/Bandwidth and CF
TRAC:IQ:SRAT 100 MHz
FREQ:CENT 500MHz
// Activate S2P file
SENSe:CORRection:FRESponse:Input1:USER:PRESet
SENSe:CORRection:FRESponse:Input1:USER:SLISt1:SELect 'c:\BP_40MHz.s2p'
SENSe:CORRection:FRESponse:Input1:USER:STATe ON
INIT:IMM; *WAI
@LOC
12.10.5
Customizing the screen layout
The remote commands required to set up the display of the R&S
FSV/A are described
here.
General display settings and items
..................................................................... 1319
..............................................................................................1323
..................................................................................... 1325
12.10.5.1
General display settings and items
The following commands add, remove or customize general display and screen ele-
ments.
Useful commands for general display settings described elsewhere
●
●
Configuring the R&S
FSV/A