Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1361
User Manual 1178.8520.02 ─ 08
RTOL
Changes of a parameter are applied from channel 2 to channel
1, but not the other way around.
<State>
ON | OFF | 0 | 1
Optional. Enables or disables coupling.
OFF | 0
Switches the coupling off
ON | 1
Switches the coupling on
*RST:
1
Return values:
<Index>
Index number of the new user-defined coupling.
Note that the returned index numbers do not necessarily have to
be the same as those shown in the user interface.
Example:
INST:COUP:USER:NEW? 'Spectrum1','All
Windows','Attenuation','Spectrum2','All
Windows','Attenuation',BID,ON
Result:
3
Synchronizes the attenuation between the channels named
'Spectrum1' and 'Spectrum2' in both directions and turns on the
coupling. Also returns the index number of the user-defined cou-
pling.
Example:
INST:COUP:USER:NEW? 'All Spectrum','All
Windows','Attenuation','','','',BID,ON
Result:
3
Synchronizes the attenuation between all Spectrum channels in
both directions and turns on the coupling. Also returns the index
number of the user-defined coupling.
Usage:
Query only
Manual operation:
See
See
INSTrument:COUPle:USER<uc>:NUMBers:LIST?
This command queries the index numbers of user-defined parameter couplings. The
index numbers are used to refer to the specific coupling in remote commands with a
USER<uc>
suffix.
Suffix:
<uc>
.
irrelevant
Configuring the R&S
FSV/A