Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
914
User Manual 1178.8520.02 ─ 08
Usage:
Query only
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>[:CACLr]:RESult?
The command returns the limit check results for the upper and lower gap (CACLR)
channels for the selected gap in an MSR ACLR measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<gap>
1 | 2
Gap (CACLR) channel number
Return values:
<LowerGap_AB>,
<UpperGap_AB>
[,<LowerGap_BC>,
<UpperGap_BC>,
<LowerGap_CD>,
<UpperGap_CD>,
<LowerGap_DE>,
<UpperGap_DE>,
<LowerGap_EF>,
<UpperGap_EF>,
<LowerGap_FG>,
<UpperGap_FG>,
<LowerGap_GH>,
<UpperGap_GH>]
Limit check results for the CACLR power in the upper and lower
gap channels for the selected gap. Results are only returned for
the available sub blocks.
PASSED
Limit check has passed.
FAIL
Limit check has failed.
NONE
No results available, e.g. because limit checking was deactiva-
ted
Example:
INIT:IMM;*WAI;
CALC:LIM:ACP:GAP2:RES?
PASSED,PASSED
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult:PHZ
<State>
This command selects the unit the R&S
FSV/A returns results for power measure-
ments.
You can query results with
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:
.
For LTE and 5G applications, use the command
FUNCtion:POWer<sb>:RESult:UNIT
.
Configuring and performing measurements