General instrument setup
R&S
®
FSVA3000/ R&S
®
FSV3000
751
User Manual 1178.8520.02 ─ 08
Exception: The SCPI Recorder maintains control of the generator even if you switch
applications.
Select
on page 751 to establish a temporary connection from the
R&S
FSV/A to the specified signal generator.
If a connection to a signal generator is already configured, the connection data is provi-
ded for information only.
Remote command:
CONFigure:GENerator:IPConnection:ADDRess
123/ABC ← IP Address or Computer name of Signal Generator
By default, the TCPIP address is expected. To enter the computer name, toggle the
"123"/"ABC" button to "ABC".
Password ← IP Address or Computer name of Signal Generator
Enter the password required to operate the connected signal generator.
Test Connection
The R&S
FSV/A attempts to establish a connection to the signal generator.
If an instrument is connected, the following information is displayed:
●
Device type
●
Name and serial number
●
Connection state
Remote command:
CONFigure:GENerator:CONNection:CSTate?
CONFigure:GENerator:CONNection[:STATe]
Connect/Disconnect
The R&S
FSV/A attempts to establish a connection to the signal generator, or discon-
nects it.
If an instrument is connected, the following information is displayed:
●
Device type
●
Name and serial number
●
Connection state
Note:
While the SCPI Recorder is enabled with generator control, you cannot discon-
nect the generator.
Remote command:
CONFigure:GENerator:CONNection[:STATe]
CONFigure:GENerator:CONNection:CSTate?
10.9
Service functions
Access
: [Setup] > "Service"
When unexpected problems arise with the R&S
FSV/A some service functions may
help you solve them.
Service functions