![Rohde & Schwarz FSV3000 User Manual Download Page 1080](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_14778431080.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1080
User Manual 1178.8520.02 ─ 08
SOURce<si>:EXTernal<gen>:ROSCillator[:SOURce]
.......................................................1080
SYSTem:COMMunicate:RDEVice:GENerator<gen>:INTerface
..........................................1080
SYSTem:COMMunicate:RDEVice:GENerator<gen>:TYPE
............................................... 1080
SYSTem:COMMunicate:TCPip:RDEVice:GENerator<gen>:ADDRess
............................... 1081
SOURce<si>:EXTernal<gen>:ROSCillator[:SOURce]
<Source>
This command controls selection of the reference oscillator for the external generator.
If the external reference oscillator is selected, the reference signal must be connected
to the rear panel of the instrument.
Suffix:
<si>
.
irrelevant
<gen>
irrelevant
Parameters:
<Source>
INTernal
Uses the internal reference.
EXTernal
Uses the external reference; if none is available, an error flag is
displayed in the status bar.
*RST:
INT
Example:
//Select an external reference oscillator
SOUR:EXT:ROSC EXT
Manual operation:
See
SYSTem:COMMunicate:RDEVice:GENerator<gen>:INTerface
<Type>
Defines the interface used for the connection to the external generator.
This command is only available if external generator control is active (see
SOURce<si>:EXTernal<gen>[:STATe]
Suffix:
<gen>
.
Parameters:
<Type>
TCPip
Example:
SYST:COMM:RDEV:GEN:INT TCP
Manual operation:
See
SYSTem:COMMunicate:RDEVice:GENerator<gen>:TYPE
<Type>
This command selects the type of external generator.
For a list of the available generator types, see
"Overview of supported generators"
Setting basic measurement parameters