![Rohde & Schwarz FSV3000 User Manual Download Page 1338](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_14778431338.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1338
User Manual 1178.8520.02 ─ 08
ATT10
Mechanical Attenuation 10 DB
ATT20
Mechanical Attenuation 20 DB
ATT40
Mechanical Attenuation 40 DB
CAL
Mechanical Calibration Source
ACDC
Mechanical Attenuation Coupling
PREamp
Preamplifier Bypass
PRES
Preselector 1: PRESEL
RFAB
Preselector 1: RFAB
PRE
Preselector 1: PREAMP30MHZ
ATT
Preselector 1: ATTINPUT2
INP
Preselector 1: INPUT2
EXT_
Preselector 2: EXT_RELAIS
Return values:
<Cycles>
Number of switching cycles.
Example:
DIAG:INFO:CCO? CAL
Usage:
Query only
Manual operation:
See
DIAGnostic:SERVice:CALibration:DATE
<CalibrationDate>
This command sets last date and time the instrument was calibrated in ISO 8601 for-
mat.
Parameters:
<CalibrationDate>
String containing calibration date of the instrument.
Example:
//Set calibration date
DIAG:SERV:CAL:DATE "2019-05-05T00:00:00Z"
Manual operation:
See
Configuring the R&S