Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1182
User Manual 1178.8520.02 ─ 08
<m>
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:X:SLIM:ZOOM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
Manual operation:
See
CALCulate<n>:THReshold
<Level>
This command defines a threshold level for the marker peak search (for
all
markers in
all
windows).
Note that you must enable the use of the threshold using
Suffix:
<n>
.
irrelevant
Parameters:
<Level>
Numeric value. The value range and unit are variable.
*RST:
-120 dBm
Default unit: DBM
Example:
CALC:THR:STAT ON
Example:
CALC:THR -82DBM
Enables the search threshold and sets the threshold value to -82
dBm.
Manual operation:
See
CALCulate<n>:THReshold:STATe
<State>
This command turns a threshold for the marker peak search on and off (for
all
markers
in
all
windows).
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
Analyzing measurements (basics)