Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1072
User Manual 1178.8520.02 ─ 08
Note that if the offset for DM-mode or CM-mode is changed, the offsets for the P-mode
and N-mode are adapted accordingly, and vice versa.
For details see the R&S
FSV/A I/Q Analyzer and I/Q Input User Manual.
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<CMOffset>
Offset of the mean voltage between the positive and negative
input terminal vs. ground
Range:
-16 V to +16 V
Default unit: V
Manual operation:
See
"Common Mode Offset / Diff. Mode Offset / P Offset / N Off-
[SENSe:]PROBe<pb>:SETup:DMOFfset
<DMOffset>
Sets the DM-mode offset. The setting is only available if a modular probe in DM-mode
is connected to the R&S
FSV/A.
If the probe is disconnected, the DM-mode offset of the probe is reset to 0.0
V.
Note that if the offset for DM-mode or CM-mode is changed, the offsets for the P-mode
and N-mode are adapted accordingly, and vice versa.
For details see the R&S
FSV/A I/Q Analyzer and I/Q Input User Manual.
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<DMOffset>
Voltage offset between the positive and negative input terminal
Default unit: V
Manual operation:
See
"Common Mode Offset / Diff. Mode Offset / P Offset / N Off-
[SENSe:]PROBe<pb>:SETup:MODE
<Mode>
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<Mode>
RSINgle | NOACtion
RSINgle
Run single: starts one data acquisition.
Setting basic measurement parameters