Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
440
User Manual 1178.8520.02 ─ 08
The used IF frequency band is indicated in the channel bar.
Problem: The analyzer does not show the signal sent by a signal generator using
the same external frontend
Possible Solution:
Although the generator and the analyzer are set to the same fre-
quency, possibly they are not using the same IF for conversion.
Make sure that both instruments are set to the same
Frontend configuration settings
.
6.2.7
Output settings
Access
: [Input/Output] > "Output"
The R&S
FSV/A can provide output to special connectors for other devices.
For details on connectors, refer to the R&S
FSV/A Getting Started manual, "Front /
Rear Panel View" chapters.
Providing trigger signals as output is described in
Chapter 6.2.10, "Trigger input/output
Providing audio output is described in
Noise Source Control
Enables or disables the 28
V voltage supply for an external noise source connected to
the "Noise source control / Power sensor") connector. By switching the supply voltage
for an external noise source on or off in the firmware, you can enable or disable the
device as required.
Data input and output