Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1128
User Manual 1178.8520.02 ─ 08
Example:
TRIG:LEV 2V
Manual operation:
See
TRIGger[:SEQuence]:LEVel:IFPower
<TriggerLevel>
This command defines the power level at the third intermediate frequency that must be
exceeded to cause a trigger event.
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed. If defined, a reference level offset is also considered.
Parameters:
<TriggerLevel>
For details on available trigger levels and trigger bandwidths,
see the data sheet.
*RST:
-20 dBm
Default unit: DBM
Example:
TRIG:LEV:IFP -30DBM
Manual operation:
See
TRIGger[:SEQuence]:LEVel:IQPower
<TriggerLevel>
This command defines the magnitude the I/Q data must exceed to cause a trigger
event.
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed. If defined, a reference level offset is also considered.
Parameters:
<TriggerLevel>
Range:
-130 dBm to 30 dBm
*RST:
-20 dBm
Default unit: DBM
Example:
TRIG:LEV:IQP -30DBM
Manual operation:
See
TRIGger[:SEQuence]:LEVel:VIDeo
<Level>
This command defines the level the video signal must exceed to cause a trigger event.
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed.
Parameters:
<Level>
Range:
0 PCT to 100 PCT
*RST:
50 PCT
Default unit: PCT
Example:
TRIG:LEV:VID 50PCT
Manual operation:
See
Setting basic measurement parameters