Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
268
User Manual 1178.8520.02 ─ 08
Child node
Attribute
Value
Parameter description
Mand.
Type
RRC | CFilter
Type of the weighting filter
Yes
Roll Off Factor
0…1
Excess bandwidth of the fil-
ter
Only if the filter
type is RRC
Bandwidth
<bandwidth in Hz>
Filter bandwidth
Only if the filter
type is RRC
FrequencyRange
Yes
Start
<frequency in Hz>
Start value of the range
Yes
Stop
<frequency in Hz>
Stop value of the range
Yes
Limit
dBm/Hz | dBm |
dBc | dBr | dB
A Range must contain
exactly two limit nodes; one
of the limit nodes has to
have a relative unit (e.g.
dBc), the other one must
have an absolute unit (e.g.
dBm)
Yes
Start
Value
<numeric_value>
Power limit at start fre-
quency
Yes
Unit
dBm/Hz | dBm |
dBc | dBr | dB
Sets the unit of the start
value
Stop
Value
<numeric_value>
Power limit at stop fre-
quency
Unit
dBm/Hz | dBm |
dBc | dBr | dB
Sets the unit of the stop
value
LimitFailMode
Absolute | Relative
| Absolute and Rel-
ative | Absolute or
Relative
If used, it has to be identical
to DefaultLimitFailMode
No
RBW
Bandwidth <bandwidth in Hz>
Yes
Type
NORM | PULS |
CFIL | RRC
No
VBW
Bandwidth <bandwidth in Hz>
Yes
Detector
NEG | POS |
SAMP | RMS |
AVER | QUAS
If used, it has to be identical
in all ranges.
No
Sweep
Mode
Manual | Auto
Yes
Time
<time in sec>
No
Amplitude
No
ReferenceLevel
Value
<power in dBm>
Yes, if the Refer-
enceLevel child
node is used
Spectrum emission mask (SEM) measurement