General instrument setup
R&S
®
FSVA3000/ R&S
®
FSV3000
743
User Manual 1178.8520.02 ─ 08
Enabling/disabling floating licenses ← License state ()
If a floating license server is configured (see
"Setup Floating License Server"
on page 744), you can enable or disable a license for the selected option in the
"License" dialog box. Two different types of licenses are available (see
You cannot change a license state that is currently in use by an application. The
change in license usage becomes effective when you select "Apply". Restarting the
firmware is not necessary.
"Enabled"
If enabled, the license is retrieved from the license server when you
select the application in the "Mode" dialog box.
Some options are available for multiple applications, such as power
sensor support. In this case, the license is retrieved if you select any
application that supports the option.
For options with suboptions, e.g. K91 with suboptions K91AC, K91N,
all suboptions are automatically activated together with the main
option by default.
"Disabled"
If the license is disabled, it is not retrieved from the server when you
open the main application. The option is grayed out in the "Ve
Options" dialog box. Disabling licenses is useful for suboptions (e.g.
K91N) or options that are available for multiple applications.
"Occupied license" period
Defines the time period an occupied license is stored locally after it is
retrieved from the license server. A time period between 1 and 7 days
is available. The license cannot be returned or disabled manually
before it expires, but it can be extended.
Remote command:
System configuration settings