Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
474
User Manual 1178.8520.02 ─ 08
●
Channel
●
RRC
●
5-Pole (not available for sweep type "FFT")
●
CISPR (6 dB) - requires EMI (R&S
FSV3-K54) option
●
MIL Std (6 dB) - requires EMI (R&S
FSV3-K54) option
For more information, see
Chapter 6.5.1.6, "Which data may pass: filter types"
Note:
The EMI-specific filter types are available if the EMI (R&S
FSV3-K54) measure-
ment option is installed, even if EMI measurement is not active. If you select a CISPR
trace detector, the filter type is automatically also set to CISPR. An active CISPR filter
is indicated in the channel info, next to the RBW information. For details, see
ter 5.13.3.1, "Resolution bandwidth and filter types"
The RBW filter configured in the bandwidth settings is identical to the filter configured
in the EMI configuration.
Remote command:
[SENSe:]BANDwidth[:RESolution]:TYPE
Default Coupling
Sets all coupled functions to the default state ("Auto"). In addition, the ratio "RBW/
VBW" is set to "Sine[1/1]" and the ratio "Span/RBW" to
100.
For more information see
Chapter 6.5.1.3, "Coupling VBW and RBW"
Remote command:
[SENSe:]BANDwidth[:RESolution]:AUTO
Sweep/Average Count
Defines the number of sweeps to be performed in the single sweep mode. Values from
0 to 200000 are allowed. If the values 0 or 1 are set, one sweep is performed.
The sweep count is applied to all the traces in all diagrams.
If the trace modes "Average", "Max Hold" or "Min Hold" are set, this value also deter-
mines the number of averaging or maximum search procedures.
In continuous sweep mode, if "Sweep Count" = 0 (default), averaging is performed
over 10 sweeps. For "Sweep Count" =1, no averaging, maxhold or minhold operations
are performed.
For more information, see
Chapter 6.5.1.8, "How much data is measured: sweep
Remote command:
Sweep Points
Defines the number of measured values to be collected during one sweep.
For details see
Chapter 6.5.1.8, "How much data is measured: sweep points and
All values from 101 to 100001 can be set. The default value is 1001 sweep points.
For EMI measurements, 200001 sweep points are available.
Bandwidth, filter and sweep configuration