Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1362
User Manual 1178.8520.02 ─ 08
Return values:
<Index>
Comma-separated list of strings
Index numbers of all available user-defined couplings
Note that the returned index numbers are not necessarily the
same as those shown in the user interface.
Example:
INST:COUP:USER:NUMB:LIST?
Result:
'1','2','4'
Number '3' is not returned, because a coupling with that index
does not exist anymore.
Usage:
Query only
Manual operation:
See
INSTrument:COUPle:USER<uc>:RELation
<Direction>
This command selects the direction in which synchronization works.
Note that the command is not available if you synchronize over all channels or all
channels of the same application.
Suffix:
<uc>
.
Index of a user-defined parameter coupling. To obtain the list of
indexes for currently defined coupling, see
Parameters:
<Direction>
LTOR | RTOL | BIDir
BIDir
Changes of a parameter are applied both ways (from channel 1
to channel 2 and vice versa).
LTOR
Changes of a parameter are applied from channel 1 to channel
2, but not the other way around.
RTOL
Changes of a parameter are applied from channel 2 to channel
1, but not the other way around.
Example:
INST:COUP:USER:REL BID
Selects bidirectional changes for the user-defined coupling with
the index number 1.
Manual operation:
See
INSTrument:COUPle:USER<uc>:REMove
[<Scope>]
This command deletes a user-defined coupling mechanism.
Configuring the R&S
FSV/A