[SENSe:]MIXer<x>:HARMonic:TYPE
<OddEven>
This command specifies whether the harmonic order to be used should be odd, even,
or both.
Which harmonics are supported depends on the mixer type.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<OddEven>
ODD | EVEN | EODD
ODD | EVEN | EODD
*RST:
EVEN
Example:
MIX:HARM:TYPE ODD
Manual operation:
See
[SENSe:]MIXer<x>:HARMonic[:LOW]
<HarmOrder>
This command specifies the harmonic order to be used for the low (first) range.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<HarmOrder>
Range:
3 to 128 (USER band); for other bands: see band
definition
*RST:
4 (for band U)
Example:
MIX:HARM 3
Manual operation:
See
[SENSe:]MIXer<x>:IF?
Queries the intermediate frequency currently used by the external mixer.
Suffix:
<x>
.
1..n
irrelevant
Example:
MIX:IF?
Example:
See
"Programming example: working with an external mixer"
Usage:
Query only