![Rohde & Schwarz FSV3000 User Manual Download Page 1115](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_14778431115.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1115
User Manual 1178.8520.02 ─ 08
[SENSe:]BWIDth[:RESolution]:FFT
<FilterMode>
[SENSe:]BANDwidth[:RESolution]:FFT
<FilterMode>
Defines the filter mode to be used for FFT filters by defining the subspan size. The
subspan is the span which is covered by one FFT analysis.
This command is only available when using the sweep type "FFT".
Parameters:
<FilterMode>
WIDE | AUTO | NARRow
AUTO
Automatically applies the sweep optimization mode that is best
for the current measurement.
NARRow
Optimizes the sweep mode for a large dynamic range.
WIDE
Optimizes the sweep mode for high performance.
*RST:
AUTO
Example:
BAND:TYPE FFT
Select FFT filter.
Example:
BAND:FFT NARR
Select narrow subspan for FFT filter.
12.7.4
Configuring the vertical axis (amplitude, scaling)
The following commands are required to configure the amplitude and vertical axis set-
tings in a remote environment.
............................................................................................... 1115
................................................................................. 1118
................................................................................... 1120
................................................................................................1122
12.7.4.1
Amplitude settings
The tasks for manual configuration are described in
Chapter 6.4.2, "Amplitude set-
Useful commands for amplitude configuration described elsewhere:
●
Remote commands exclusive to amplitude configuration:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
..........................................................1116
.......................................................................................... 1116
Setting basic measurement parameters