Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1173
User Manual 1178.8520.02 ─ 08
CALCulate<n>:DELTamarker<m>:MODE
<Mode>
This command defines whether the position of a delta marker is provided as an abso-
lute value or relative to a reference marker. Note that this setting applies to
all
win-
dows.
Note that when the position of a delta marker is
queried
, the result is always an abso-
lute value (see
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<Mode>
ABSolute
Delta marker position in absolute terms.
RELative
Delta marker position in relation to a reference marker.
*RST:
RELative
Example:
CALC:DELT:MODE ABS
Absolute delta marker position.
CALCulate<n>:DELTamarker<m>:MREFerence
<Reference>
This command selects a reference marker for a delta marker other than marker 1.
Suffix:
<n>
.
<m>
Parameters:
<Reference>
Example:
CALC:DELT3:MREF 2
Specifies that the values of delta marker 3 are relative to marker
2.
Manual operation:
See
CALCulate<n>:DELTamarker<m>[:STATe]
<State>
This command turns delta markers on and off.
If necessary, the command activates the delta marker first.
No suffix at DELTamarker turns on delta marker 1.
Suffix:
<n>
.
Analyzing measurements (basics)