Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1119
User Manual 1178.8520.02 ─ 08
Manual operation:
See
INPut<ip>:ATTenuation:AUTO:MODE
<OptMode>
Selects the priority for signal processing
after
the RF attenuation has been applied.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<OptMode>
LNOise | LDIStortion
LNOise
Optimized for high sensitivity and low noise levels
LDIStortion
Optimized for low distortion by avoiding intermodulation
*RST:
LDIStortion (WLAN application: LNOise)
Example:
INP:ATT:AUTO:MODE LNO
INPut<ip>:EATT
<Attenuation>
This command defines an electronic attenuation manually. Automatic mode must be
switched off (
INP:EATT:AUTO OFF
, see
If the current reference level is not compatible with an attenuation that has been set
manually, the command also adjusts the reference level.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Attenuation>
attenuation in dB
Range:
see data sheet
Increment: 1 dB
*RST:
0 dB (OFF)
Default unit: DB
Example:
INP:EATT:AUTO OFF
INP:EATT 10 dB
Manual operation:
See
"Using Electronic Attenuation"
INPut<ip>:EATT:AUTO
<State>
This command turns automatic selection of the electronic attenuation on and off.
If on, electronic attenuation reduces the mechanical attenuation whenever possible.