Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1055
User Manual 1178.8520.02 ─ 08
12.7.1.3
Using external mixers
The commands required to work with external mixers in a remote environment are
described here. Note that these commands require the R&S
FSV/A to have an external
mixer option installed and an external mixer to be connected to the R&S
FSV/A.
For details on working with external mixers see
Chapter 6.2.5.1, "Basics on external
Conversion loss table settings
............................................................................ 1064
Programming example: working with an external mixer
......................................1068
Basic settings
The basic settings concern general usage of an external mixer.
...........................................................................................1055
......................................................................................1055
.....................................................................................1056
........................................................................................ 1056
............................................................................................1056
......................................................................................1057
[SENSe:]MIXer<x>[:STATe]
<State>
Activates or deactivates the use of a connected external mixer as input for the mea-
surement. This command is only available if the optional External Mixer is installed and
an external mixer is connected.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
MIX ON
Manual operation:
See
[SENSe:]MIXer<x>:BIAS:HIGH
<BiasSetting>
This command defines the bias current for the high (last) range.
This command is only available if the external mixer is active (see
Suffix:
<x>
.
1..n
irrelevant
Setting basic measurement parameters