Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1137
User Manual 1178.8520.02 ─ 08
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
●
Center frequency
●
Reference level
Example:
ADJ:ALL
Manual operation:
See
"Adjusting all Determinable Settings Automatically (Auto
[SENSe:]ADJust:CONFigure:LEVel:DURation
<Duration>
To determine the ideal reference level, the R&S
FSV/A performs a measurement on
the current input data. This command defines the length of the measurement if
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
is set to
MANual
.
Parameters:
<Duration>
Numeric value in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Manual operation:
See
"Changing the Automatic Measurement Time (Meastime
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
<Mode>
To determine the ideal reference level, the R&S
FSV/A performs a measurement on
the current input data. This command selects the way the R&S
FSV/A determines the
length of the measurement .
Parameters:
<Mode>
AUTO
The R&S
FSV/A determines the measurement length automati-
cally according to the current input data.
MANual
The R&S
FSV/A uses the measurement length defined by
[SENSe:]ADJust:CONFigure:LEVel:DURation
on page 1137.
*RST:
AUTO
Setting basic measurement parameters