Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1021
User Manual 1178.8520.02 ─ 08
●
Instrument setup and querying of the result list at the end of the measurement:
With this method, the control computer may be used for other activities while the
measurement is being performed. However, more time is needed for synchroniza-
tion via service request.
12.5.15.1
Performing pulse power measurements
The following commands control pulse power measurements.
CALCulate<n>:MARKer<m>:FUNCtion:MSUMmary
........................................................ 1021
.............................................................................................1022
................................................................................. 1022
..................................................................................... 1022
.................................................................................... 1024
CALCulate<n>:MARKer<m>:FUNCtion:MSUMmary
<TimeOffset>, <MeasTime>,
<PulsePeriod>, <OfPulses>
This command configures power measurements on pulses in the time domain.
To evaluate the pulse power, the R&S
FSV/A uses the data captured during a previous
measurement. The data recorded during the set measurement time is combined to a
measured value for each pulse according to the detector specified and the indicated
number of results is output as a list.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Parameters:
<TimeOffset>
Defines a time offset to start the measurement at the first pulse
of a trace.
*RST:
0
Default unit: s
<MeasTime>
Defines the measurement time.
Default unit: s
Configuring and performing measurements