Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
894
User Manual 1178.8520.02 ─ 08
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:LIM:ACP:GAP1:ACLR:REL:STAT ON
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>[:AUTO][:CACLr][:RELative]
<Limit>, <UpperLimit>
This command defines the relative limit of the specified gap (CACLR) channel. The ref-
erence value for the relative limit is the measured channel power.
If you define both an absolute limit and a relative limit, the R&S
FSV/A uses the lower
value for the limit check.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<gap>
1 | 2
Gap (CACLR) channel number
Parameters:
<Limit>
Defines the relative limit of the specified gap channel in dB.
Default unit: DB
<UpperLimit>
Optional: Defines the relative upper limit of the specified gap
channel.
Default unit: dB
Example:
CALC:LIM:ACP:GAP1:REL 3dB,0
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>[:AUTO][:CACLr][:RELative]:STATe
<State>
This command turns the relative limit check for the specified gap (CACLR) channel on
and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe]
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<gap>
1 | 2
Gap (CACLR) channel number
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
Configuring and performing measurements