Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
908
User Manual 1178.8520.02 ─ 08
Parameters for setting and query:
<SBGaps>
AB | BC | CD | DE | EF | FG | GH
Name of the gap, defined by the letters of the surrounding sub
blocks (e.g. "AB" for the gap between sub blocks A and B).
Example:
SENS:POW:ACH:FILT-ALPH:GAP:MAN:LOW BC,0.25
Example:
See
"Example: configuring and performing an asymmetrical
Manual operation:
See
[SENSe:]POWer:ACHannel:FILTer:ALPHa:GAP<gap>:MANual:UPPer
<SBGaps>,
<Alpha>
[SENSe:]POWer:ACHannel:FILTer:ALPHa:GAP<gap>:MANual:UPPer?
<SBGaps>
This command defines the roll-off factor for the specified upper gap channel's weight-
ing filter.
This command is only available for for asymmetrical (manual) configuration of gap
channels (see
[SENSe:]POWer:ACHannel:GAP<gap>:MODE
Suffix:
<gap>
.
1 | 2
Gap channel number
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Parameters for setting and query:
<SBGaps>
AB | BC | CD | DE | EF | FG | GH
Name of the gap, defined by the letters of the surrounding sub
blocks (e.g. "AB" for the gap between sub blocks A and B).
Example:
SENS:POW:ACH:FILT:ALPH:GAP2:MAN:UPP BC,0.25
Example:
See
"Example: configuring and performing an asymmetrical
Manual operation:
See
[SENSe:]POWer:ACHannel:GCHannel[:STATe]:GAP<gap>:MANual:LOWer
<SBGaps>, <State>
[SENSe:]POWer:ACHannel:GCHannel[:STATe]:GAP<gap>:MANual:LOWer?
<SBGaps>
Defines which lower gap channels are active in the specified gap.