Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
872
User Manual 1178.8520.02 ─ 08
Parameters:
<Number>
Range:
1 to 18
*RST:
1
Manual operation:
See
12.5.3.3
Defining weighting filters
The following commands define weighting filters for ACLR measurements.
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel
........................................................ 872
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL]
............................................................... 872
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<ch>
................................................. 872
[SENSe:]POWer:ACHannel:FILTer:ALPHa:CHANnel<ch>
.................................................. 873
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ACHannel
....................................................... 873
[SENSe:]POWer:ACHannel:FILTer[:STATe][:ALL]
.............................................................. 873
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALTernate<ch>
................................................ 873
[SENSe:]POWer:ACHannel:FILTer[:STATe]:CHANnel<ch>
................................................. 874
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel
<Alpha>
This command defines the roll-off factor for the adjacent channel weighting filter.
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Manual operation:
See
See
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL]
<Value>
This command defines the alpha value for the weighting filter for all channels.
Parameters:
<Value>
*RST:
0.22
Example:
POW:ACH:FILT:ALPH:ALL 0.35
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<ch>
<Alpha>
This command defines the roll-off factor for the alternate channel weighting filter.
Suffix:
<ch>
.
1..n
Alternate channel number
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Configuring and performing measurements