Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
984
User Manual 1178.8520.02 ─ 08
Parameters:
<Magnitude>
The number is a statistical value and therefore dimensionless.
Range:
1E-5 to 1.0
*RST:
1.0
Example:
CALC:STAT:SCAL:Y:UPP 0.01
Manual operation:
See
12.5.8.5
Performing a statistical measurement
The following commands are required to perform a statistical measurement:
Chapter 12.5.1, "Performing mea-
12.5.8.6
Retrieving results
The following commands are required to retrieve the measurement results.
Useful commands for retrieving results described elsewhere:
●
Remote commands exclusive to statistical results
CALCulate<n>:STATistics:CCDF:X<t>?
............................................................................984
CALCulate<n>:STATistics:RESult<res>?
.......................................................................... 985
CALCulate<n>:STATistics:CCDF:X<t>?
<Probability>
This command queries the results of the CCDF.
Suffix:
<n>
.
<t>
Query parameters:
<Probability>
P0_01
Level value for 0.01 % probability
P0_1
Level value for 0.1 % probability
P1
P1: Level value for 1 % probability
P10
Level value for 10 % probability
Return values:
<CCDF Result>
Configuring and performing measurements