Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1177
User Manual 1178.8520.02 ─ 08
Suffix:
<n>
.
<m>
Parameters:
<Trace>
1 to 6
Trace number the marker is assigned to.
Example:
//Assign marker to trace 1
CALC:MARK3:TRAC 2
Manual operation:
See
"Assigning the Marker to a Trace"
CALCulate<n>:MARKer<m>:X
<Position>
This command moves a marker to a specific coordinate on the x-axis.
If necessary, the command activates the marker.
If the marker has been used as a delta marker, the command turns it into a normal
marker.
Suffix:
<n>
.
<m>
Parameters:
<Position>
Numeric value that defines the marker position on the x-axis.
The unit depends on the result display.
Range:
The range depends on the current x-axis range.
Default unit: Hz
Example:
CALC:MARK2:X 1.7MHz
Positions marker 2 to frequency 1.7 MHz.
Manual operation:
See
"Marker 1/Marker 2/Marker 3/Marker 4"
See
See
See
See
12.8.3.2
General marker settings
The following commands control general marker functionality.
Remote commands exclusive to general marker functionality
......................................................................................1178
DISPlay[:WINDow<n>]:MINFo[:STATe]
........................................................................... 1178
CALCulate<n>:MARKer<m>:X:SSIZe
.............................................................................1178
Analyzing measurements (basics)