
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
880
User Manual 1178.8520.02 ─ 08
<UpperLimit>
The limit of the upper alternate channel.
Range:
0 dB to 100 dB
*RST:
0 dB
Default unit: dB
Manual operation:
See
See
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:RESult?
This command queries the state of the limit check for the adjacent or alternate chan-
nels in an ACLR measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<ch>
Alternate channel number
Return values:
<LowerChan>
text value
The state of the limit check for the lower alternate or adjacent
channels.
PASSED
Limit check has passed.
FAIL
Limit check has failed.
<UpperChan>
text value
The state of the limit check for the upper alternate or adjacent
channels.
PASSED
Limit check has passed.
FAIL
Limit check has failed.
Example:
INIT:IMM;*WAI;
CALC:LIM:ACP:ACH:RES?
PASSED,PASSED
Usage:
Query only
Configuring and performing measurements