Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
276
User Manual 1178.8520.02 ─ 08
Remote command:
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STARt
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STOP
Filter Type
Sets the filter type for this range.
For details on filter types, see
Chapter 6.5.1.6, "Which data may pass: filter types"
Remote command:
[SENSe:]LIST:RANGe<ri>:FILTer:TYPE
RBW
Sets the RBW value for this range.
For details on the RBW, see
Chapter 6.5.1.1, "Separating signals by selecting an
appropriate resolution bandwidth"
Remote command:
[SENSe:]LIST:RANGe<ri>:BANDwidth:RESolution
VBW
Sets the VBW value for this range.
For details on the VBW, see
Chapter 6.5.1.2, "Smoothing the trace using the video
Remote command:
[SENSe:]LIST:RANGe<ri>:BANDwidth:VIDeo
Sweep Time Mode
Activates or deactivates the auto mode for the sweep time.
For details on the sweep time mode, see
Chapter 6.5.1.7, "How long the data is mea-
Remote command:
[SENSe:]LIST:RANGe<ri>:SWEep:TIME:AUTO
Sweep Time
Sets the sweep time value for the range.
For details on the sweep time, see
Chapter 6.5.1.7, "How long the data is measured:
Remote command:
[SENSe:]LIST:RANGe<ri>:SWEep:TIME
Detector
Sets the detector for the range.
"Mapping samples to sweep points with the trace detector"
Remote command:
[SENSe:]LIST:RANGe<ri>:DETector
Spurious emissions measurement