Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1292
User Manual 1178.8520.02 ─ 08
<ext>
1..n
Example:
ROSC:EXT:FREQ 13MHZ
Sets the frequency to 13 MHz.
SOUR:EXT:ROSC:EXT:FREQ 13MHZ
Manual operation:
See
[SENSe:]ROSCillator:SOURce
<Source>
This command selects the reference oscillator.
If you want to select the external reference, it must be connected to the R&S
FSV/A.
Parameters:
<Source>
INTernal
The internal reference is used (10 MHz).
EXTernal | EXTernal1 | EXT1
The external reference from the "REF INPUT 10
MHZ" connec-
tor is used; if none is available, an error flag is displayed in the
status bar.
EXTernal2 EXT2
The external reference from the "REF Input 2" connector is used
(requires option R&S
FSV3-K703); if none is available, an error
flag is displayed in the status bar.
EAUTo
The external reference is used as long as it is available, then the
instrument switches to the internal reference.
Example:
ROSC:SOUR EXT
Manual operation:
See
[SENSe:]ROSCillator:SOURce:EAUTo?
This command queries the current reference type in case you have activated an auto-
matic switch to the internal reference if the external reference is missing.
Return values:
<Reference>
INT | EXT
INT
internal reference
EXT
external reference
Example:
SENS:ROSC:SOUR:EAUT?
Queries the currently available reference type.
Usage:
Query only