![Rohde & Schwarz FSV3000 User Manual Download Page 1351](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_14778431351.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1351
User Manual 1178.8520.02 ─ 08
INSTrument:COUPle:ATTen
<State>
This command turns synchronization of the attenuation and unit between measure-
ment channels on and off.
Parameters:
<State>
ALL | NONE
ALL
Turns on synchronization.
NONE
Turns off synchronization.
*RST:
ALL
Example:
INST:COUP:ATT ALL
Synchronizes the attenuation.
Manual operation:
See
"Synchronizing parameters across all measurement chan-
INSTrument:COUPle:AUNit
<State>
This command turns synchronization of the amplitude unit configuration between mea-
surement channels on and off.
Parameters:
<State>
ALL | NONE
ALL
Turns on synchronization.
NONE
Turns off synchronization.
*RST:
NONE
Example:
INST:COUP:AUN ALL
Manual operation:
See
"Synchronizing parameters across all measurement chan-
INSTrument:COUPle:BANDwidth
<State>
INSTrument:COUPle:BWIDth
<State>
This command turns synchronization of the resolution bandwidth (and filter type)
between measurement channels on and off.
Parameters:
<State>
ALL | NONE
ALL
Turns on synchronization.
NONE
Turns on synchronization.
*RST:
NONE
Configuring the R&S
FSV/A