Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
243
User Manual 1178.8520.02 ─ 08
For details on the sweep time mode, see
Chapter 6.5.1.7, "How long the data is mea-
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:SWEep:TIME:AUTO
Sweep Time
Sets the sweep time value for the range.
For details on the sweep time, see
Chapter 6.5.1.7, "How long the data is measured:
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:SWEep:TIME
Ref Level
Sets the reference level for the range.
For details on the reference level, see
Chapter 6.4.1.1, "Reference level"
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:RLEVel
RF Att Mode
Activates or deactivates the auto mode for RF attenuation.
For details on attenuation, see
Chapter 6.4.1.2, "RF attenuation"
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:ATTenuation:AUTO
RF Attenuation
Sets the attenuation value for the range.
For details on attenuation, see
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:ATTenuation
Preamp
Switches the preamplifier on or off.
For details on the preamplifier, see
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:GAIN:STATe
Transducer Factor
Sets a transducer for the specified range. You can only choose a transducer that fulfills
the following conditions:
●
The transducer overlaps or equals the span of the range.
●
The x-axis is linear.
●
The unit is dB.
For details on transducers, see
Chapter 10.5.1, "Basics on transducer factors"
Spectrum emission mask (SEM) measurement