Amplitude and vertical axis configuration
In the Spectrum application, measurement results usually consist of the measured sig-
nal levels (amplitudes) displayed on the vertical (y-)axis for the determined frequency
spectrum or for the measurement time (horizontal, x-axis). The settings for the vertical
axis, regarding amplitude and scaling, are described here.
Impact of the vertical axis settings
........................................................................ 453
.................................................................................................455
How to optimize the amplitude display
..................................................................462
6.4.1
Impact of the vertical axis settings
Some background knowledge on the impact of the described settings is provided here
for a better understanding of the required configuration.
6.4.1.1
Reference level
The reference level value is the maximum value the AD converter can handle without
distortion of the measured value. Signal levels above this value will not be measured
correctly, which is indicated by the "IF Overload" status display.
Internally, the reference level is also used to determine the optimum hardware settings
for the R&S
FSV/A. The defined reference level should correspond with the maximum
expected RF input level.
When determining the expected input level, consider that the power from
all
input sig-
nals contribute to the total power. The reference level must be higher than the total
power from all signals.
The optimum reference level for the current measurement settings can be set automat-
ically by the R&S
FSV/A (see
The reference level determines the amplitude represented by the topmost grid line in
the display. When you change the reference level, the measurement is not restarted;
the results are merely shifted in the display. Only if the reference level changes due to
a coupled RF attenuation (see
ment is restarted.
In general, the R&S
FSV/A measures the signal voltage at the RF input. The level dis-
play is calibrated in RMS values of an unmodulated sine wave signal. In the default
state, the level is displayed at a power of 1 mW (= dBm). Via the known input impe-
dance, conversion to other units is possible.