Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
971
User Manual 1178.8520.02 ─ 08
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<SweepTime>
Sweep time.
The range depends on the ratios of the span to the RBW and
the RBW to the VBW. Refer to the data sheet for more informa-
tion.
Example:
LIST:RANG2:SWE:TIME 2MS
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:SWEep:TIME:AUTO
<State>
This command turns automatic selection of the sweep time for a spurious emission
measurement range on and off.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
LIST:RANG2:SWE:TIME:AUTO ON
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:TRANsducer
<Transducer>
This command selects a transducer factor for a spurious emission measurement
range.
Note the following prerequisites for the selected transducer:
●
The transducer must cover at least the span of the range.
●
The x-axis has to be linear.
●
The unit has to be dB.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Transducer>
String containing the transducer file name. Do not include a file
extension or the file path.
The file must be located in the
C:\R_S\INSTR\trd
directory.
Example:
LIST:RANG2:TRAN 'MYTRANS'
Manual operation:
See
Configuring and performing measurements