Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1003
User Manual 1178.8520.02 ─ 08
Return values:
<TOI>
Maximum third order intercept point.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:TOI ON
Switches the intercept measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:TOI:RES:MAX?
Returns the maximum TOI.
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult:MINimum?
This command queries the results for the minimum third order intercept point measure-
ment (see
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<TOI>
Minimum third order intercept point.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:TOI ON
Switches the intercept measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:TOI:RES:MIN?
Returns the minimum TOI.
Usage:
Query only
12.5.11.2
Programming example: measuring the TOI
This example demonstrates how to determine the TOI in a remote environment.
//-----------Configuring the measurement ------------
*RST
//Reset the instrument
CALC:MARK:FUNC:TOI ON
//Activate TOI measurement.
Configuring and performing measurements