Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
153
User Manual 1178.8520.02 ─ 08
Measuring the edges of a GSM burst with high time resolution
Due to the high time resolution of the R&S
FSV/A at the 0 Hz display range, the edges
of TDMA bursts can be measured precisely. The edges can be shifted to the display
area using the trigger offset.
Figure 5-15: Test setup
Table 5-6: Signal generator settings (e.g. R&S SMW)
Frequency
890 MHz
Level
0 dBm
Modulation
GSM, one timeslot activated
The measurement is based on the example
"Measuring the power of a GSM burst dur-
1. Switch off the power measurement.
a) Press the [MEAS] key.
b) Select "Zero Span".
2. Increase the time resolution by setting the sweep time to
100
µs
([SWEEP] >
"Sweep Time Manual" ).
3. Shift the rising edge of the GSM burst to the center of the display by defining a trig-
ger offset.
a) Press the [TRIG] key.
b) Select "Trigger Offset".