Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1081
User Manual 1178.8520.02 ─ 08
Suffix:
<gen>
.
Parameters:
<Name>
<Generator name as string value>
*RST:
SMU02
Example:
//Select an external generator
SYST:COMM:RDEV:GEN:TYPE 'SMW06'
Manual operation:
See
SYSTem:COMMunicate:TCPip:RDEVice:GENerator<gen>:ADDRess
<Address>
Configures the TCP/IP address for the external generator.
Suffix:
<gen>
.
Parameters:
<Address>
TCP/IP address between 0.0.0.0 and 0.255.255.255
*RST:
0.0.0.0
Example:
SYST:COMM:TCP:RDEV:GEN:ADDR 130.094.122.195
Manual operation:
See
"TCPIP Address / Computer Name"
Source calibration
The following commands are required to activate the calibration functions of the exter-
nal tracking generator. However, they are only available if external generator control is
active (see
SOURce<si>:EXTernal<gen>[:STATe]
Useful commands for source calibration described elsewhere:
●
Chapter 12.10.3, "Working with transducers"
●
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:
RPOSition
Remote commands exclusive to source calibration:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RVALue
..........................1081
[SENSe:]CORRection:COLLect[:ACQuire]
......................................................................1082
..................................................................................... 1083
........................................................................................1083
.......................................................................................1083
[SENSe:]CORRection:TRANsducer:GENerate
................................................................ 1084
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RVALue
<Value>
This command defines the reference value assigned to the reference position in the
specified window. Separate reference values are maintained for the various displays.
Setting basic measurement parameters