![Rohde & Schwarz FSV3000 User Manual Download Page 1006](http://html.mh-extra.com/html/rohde-and-schwarz/fsv3000/fsv3000_user-manual_14778431006.webp)
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1006
User Manual 1178.8520.02 ─ 08
//Activate AM modulation depth measurement.
//--------------Performing the Measurement-----
INIT:CONT OFF
//Selects single sweep mode.
INIT:IMM
// Perform a single measurement
CALC:MARK:FUNC:MDEP:SEAR ONCE
//Initiates a search for signals in the current trace.
//---------------Retrieving Results-------------
CALC:MARK:FUNC:MDEP:RES?
//Queries the measurement results.
//If the results are not accurate, change the position of the
//the temporary markers manually.
//----Changing the position of the temp markers-----
CALC:MARK:X 100MHZ
//Positions the reference marker on 100 MHz.
CALC:DELT2:X 5KHZ
//Positions delta marker 2 and 3 at a distance of 5 kHz to the reference marker.
CALC:DELT3:X 1KHZ
//Corrects the position of delta marker 3 by 1 kHz.
CALC:MARK:FUNC:MDEP:RES?
//Queries the measurement results for the repositioned markers.
12.5.13
Remote commands for EMI measurements
The following commands are required to perform EMI measurements in a remote envi-
ronment. This measurement requires the R&S
FSV3-K54 option.
The following tasks specific to the EMI application are described here:
.............................................................................. 1006
.................................................................................... 1007
Configuring the EMI final test
..............................................................................1008
...................................................................................1009
......................................................................................... 1011
..........................................................................................1013
Programming example: EMI measurement
.........................................................1013
12.5.13.1
Activating EMI measurement
EMI measurement must be activated explicitely.
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement[:STATe]
................................ 1007
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement[:STATe]
....................................... 1007
Configuring and performing measurements