General instrument setup
R&S
®
FSVA3000/ R&S
®
FSV3000
753
User Manual 1178.8520.02 ─ 08
This data is stored to the C:\R_S\INSTR\USER directory on the instrument.
The file name consists of the unique device ID and the current date and time of the file
creation.
If you contact the Rohde
&
Schwarz support to get help for a certain problem, send
these files to the support in order to identify and solve the problem faster.
Remote command:
Save Device Footprint
Creates an
*.xml
file with information on installed hardware, software, image and
FPGA versions. The
*.xml
file is stored under
C:\Program Files\Rohde-Schwarz\FSV3000\<version>\devicedata\xml\
on the instrument. It is also included in the service ZIP file (see
Remote command:
Last Service Date
Opens a calendar to define the date that the R&S
FSV/A was inspected by
Rohde
&
Schwarz service. The service date is updated by the service technicians. For
new instruments, the last service date (= date of production) is entered during produc-
tion.
If the field contains no date, we recommend that you enter the date of last service
according to the service certificates you received after the R&S
FSV/A has been serv-
iced.
If the R&S
FSV/A has not been serviced yet, leave the field empty.
Remote command:
Last Calibration Date
Opens a calendar to define the date that the R&S
FSV/A was calibrated. For new
instruments, we recommend to enter the factory calibration date specified on the cali-
bration certificate.
Remote command:
DIAGnostic:SERVice:CALibration:DATE
Next Calibration Due
Opens a calendar to define the date that the R&S
FSV/A needs its next calibration.
The recommended calibration interval is specified in a read only field next to this input
field.
This date must always be configured by the customer due to ISO17025.
Remote command:
Set date:
DIAGnostic:SERVice:CALibration:DUE:DATE
Recommended interval:
DIAGnostic:SERVice:CALibration:INTerval?
Service functions