Common analysis and display functions
R&S
®
FSVA3000/ R&S
®
FSV3000
577
User Manual 1178.8520.02 ─ 08
File contents
Description
XAxisUnit;FREQ_HZ
Unit of x values
XAxisScaleMode;ABSOLUTE
Scaling of x-axis (absolute or relative)
YAxisUnit;LEVEL_DB
Unit of y values
YAxisScaleMode;ABSOLUTE
Scaling of y-axis (absolute or relative)
NoOfPoints
;5
Number of points the line is defined by
Data section for individual data points
-4500000000;-50
x- and y-values of each data point defining the line
-2000000000;-30
-1000000000;0
0;-30
2500000000;-50
7.5
Trace configuration
A trace is a collection of measured data points. The trace settings determine how the
measured data is analyzed and displayed on the screen.
7.5.1
Standard traces
7.5.1.1
Basics on setting up traces
Some background knowledge on traces is provided here for a better understanding of
the required configuration settings.
Mapping samples to sweep points with the trace detector
................................... 577
....................................................................................581
Analyzing several traces - trace mode
..................................................................582
How many traces are averaged - sweep count + Sweep mode
............................583
How trace data is averaged - the averaging mode
............................................... 584
Mapping samples to sweep points with the trace detector
A trace displays the values measured at the sweep points. The number of samples
taken during a sweep can be much larger than the number of sweep points that are
displayed in the measurement trace.
Trace configuration