CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:LIMit<li>:LDELta?
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:LIMit<li>:LDELta?
<Amplitude>
This command queries the vertical distance from the marker position to the limit line.
The unit is dB.
If the marker has been assigned to a different trace than the limit line, or if no limit ine
is defined for the marker position, the command returns -200.
Suffix:
<n>
.
<m>
<li>
1..n
Return values:
<Amplitude>
Vertical distance to the limit line in dB.
Example:
CALC:MARK3:FUNC:FME:LIM2:LDEL?
Queries the distance of marker 3 to the second limit line.
Usage:
Query only
12.5.13.7
Evaluating the results
The commands required to control the demodulation of signals at the marker position
are described in
Chapter 12.8.3.14, "Marker demodulation"
12.5.13.8
Programming example: EMI measurement
This example demonstrates how to detect electromagnetic interferences (EMI) in a
remote environment.
//----------- Preparing the measurement ------------
//Reset the instrument
*RST
//Define the span to be analyzed
FREQ:STAR 150kHz
FREQ:STOP 1GHz
//Configure two traces, one with peak detector, one with average detector
DISP:TRAC1 ON
DISP:TRAC2 ON
DET1 POS
DET2 AVER
//----------- Configuring the measurement ------------
//Select EMI measurement
CALC:MARK:FUNC:FME:STAT ON
//Configure CISPR filter and RBW
Configuring and performing measurements